5 Reasons Why Your Business Needs Bell-Prover Gas Meter Test Bench?

31 Mar.,2025

 

GS-ENG-09-01—Requirements for the certification by correlation of ...

Table of contents

  • 1.0 Scope
  • 2.0 Authority
  • 3.0 Terminology
  • 4.0 Administrative requirements
    • 4.1 General
    • 4.2 Gas measuring apparatus requirements
    • 4.3 Statistics
    • 4.4 Certification testing
    • 4.5 Roles and responsibilities
      • 4.5.1 Designating authority
      • 4.5.2 Owner
    • 4.6 Statement of intended use of the gas measuring apparatus
      • 4.6.1 Limitations
      • 4.6.2 Statement of intended use - Details
  • 5.0 Metrological requirements
    • 5.1 Environment
      • 5.1.1 Temperature
    • 5.2 Prover oil
    • 5.3 Mechanical requirements
      • 5.3.1 Bell component installation
      • 5.3.2 Prover oil level
      • 5.3.3 Bell balance
      • 5.3.4 Bell reference pressure
      • 5.3.5 Bell static pressure
      • 5.3.6 Bell dynamic pressure
      • 5.3.7 Leak test
        • 5.3.7.1 System leak test
        • 5.3.7.2 Operational leak test
      • 5.3.8 Flow rate tests
      • 5.3.9 Register verification
    • 5.4 Meter classifications and transfer meters
      • 5.4.1 Meter classifications
      • 5.4.2 Transfer meters
    • 5.5 Volume correlations
      • 5.5.1 Direct counting gas measuring apparatus
      • 5.5.2 Inferential gas measuring apparatus
      • 5.5.3 Correlations
      • 5.5.4 Maximum error detection
      • 5.5.5 Additional modes of operation
  • 6.0 Technical requirements
    • 6.1 Use requirements
      • 6.1.1 Weekly correlation - Transfer meter / local volumetric standard
      • 6.1.2 Daily correlation - Transfer meter / gas measuring apparatus
      • 6.1.3 Operational leak detection
      • 6.1.4 Prover oil
      • 6.1.5 Temperature
      • 6.1.6 Maintenance
Record of changes Version Date Description 1.0 -07-06
  • Removed the unit for relative density in section 5.2.1.
  • Added a reference for the American Society for Testing Materials test method specifications in section 5.2.5.
  • Corrected the tolerance for pressure in sections 5.3.4.2, 5.3.5.2 and 5.3.6.2.
  • Changed the name of the senior engineer.
  • Corrected spelling and formatting.

1.0 Scope

1.1 This document outlines the requirements for the certification, recertification, calibration and use of working level gas measuring apparatus using bell provers up to 10 ft³ capacity. These measuring apparatus are used for the verification, reverification and/or compliance sampling of gas meters. Bell provers with a volumetric capacity greater than 10 ft³ must be processed for certification on an individual basis as determined by Measurement Canada.

For more information, please visit TNMA.

1.2 This document is supported by the procedures set out in GS-ENG-09-01.1—Procedures and Worksheets for Calibrating and Certifying Gas Measuring Apparatus—Working Level Bell Provers pursuant to the Requirements of GS-ENG-09-01.

1.3 This certification document is considered as interim, as it does not fully incorporate the requirements of S-S-02—Measurement Uncertainty and Meter Conformity Evaluation Specifications. However, the determination of measurement uncertainty is subject to the guidelines in GS-ENG-09-04—Guidelines for the Determination of Measurement Uncertainty in Working Level Bell Provers—Correlation Method.

2.0 Authority

This document is issued pursuant to sections 7 and 8 of the Electricity and Gas Inspection Regulations. It has been produced under the delegated authority of the Senior Engineer, Gas Measurement, Measurement Canada for the purposes of setting out the requirements for the calibration, certification and use of gas measuring apparatus.

Christian Lachance, P. Eng.
Senior Engineer ‑ Natural Gas Measurement
Engineering and Laboratory Services Directorate
Measurement Canada

3.0 Terminology

Calibration:
Comparison between two instruments, measuring apparatus or standards, one of which is of known accuracy. Tests are performed to detect, correlate, report or eliminate by adjustment any variation in accuracy of the instrument or measuring apparatus of unknown accuracy.
Cardinal point:
A major volume increment marking on the bell scale chosen as a point of reference for a selected test volume.
Certification:
A process which ensures that a measuring apparatus has been properly calibrated and installed for its intended use, and that an acceptable accuracy correlation exists between it and a reference standard.
Certification testing:
A specialized form of calibration performed according to fixed standards which must be met prior to the issuance of the certification certificate.
Designating authority:
Individual delegated the authority to certify gas measuring apparatus under the Electricity and Gas Inspection Act and associated Measurement Canada policies.
Direct counting gas measuring apparatus:
A gas measuring apparatus which determines the meter error using register revolutions of the meter under test.
High load rate:
The flow rate corresponding to the specified high load test flow rate, i.e. 145 % ± 5 % of a diaphragm meter's rated capacity of air at 0.5 inches differential pressure.
Inferential gas measuring apparatus:
A gas measuring apparatus which determines meter error by a method other than direct counting.
Initial certification:
Certification of gas measuring apparatus for the first time.
Local volumetric standard:
A master bell prover or certified transfer prover located at or near the site of the gas measuring apparatus.
Low load flow rate:
The term used to describe the flow rate corresponding to the specified low load test flow rate, i.e. 45 % ± 5 % of a diaphragm meter's rated capacity of air at 0.5 inches differential pressure.
Master bell prover:
The local volumetric standard which is traceable to a national volumetric reference standard.
Meter class:
A general grouping of meter types having varied manufacturers and model designations, but having the same units of measure and similar rated capacities of air at 0.5 inches differential pressure. Class designations (shown in cubic feet/hour): 100 class (<140), 200 class (140 to 200), 300 class (201 to 300), 400 class (301 to 350), 500 class (351 to 450), 600 class (451 to 500), 700 class (501 to 550), 800 class (551 to 650), 900 class (651 to 700), class (701 to 800). All other meters must be grouped into classes based on 99 ft3 intervals or the International System of Units equivalent.
Meter classification:
A grouping of meters having the same manufacturer, meter class, and units of measure, formed from the listing of meters in the owner's statement of intended use.
Minimum range volume:
The nominal volume range over which a gas measuring apparatus is to be calibrated.
Monitor:
To observe, record or detect an operation or condition with instruments.
Non-converting meter:
A meter that does not correct the registered volumes for pressure and/or temperature.
Owner:
The owner of the gas measuring apparatus to be calibrated and certified or recertified.
Recertification:
Certification of a gas measuring apparatus subsequent to the initial certification.
Relative error:
The absolute error of measurement divided by the true (conventional) value of the measurand. The measurand is a quantity subjected to measurement.
Start point:
The fully raised position of the bell prior to commencement of the test sequence.
Technical evaluator (inspector):
Individual appointed by the designating authority and delegated under the Electricity and Gas Inspection Act and associated Measurement Canada policies to perform certification testing of a gas measuring apparatus.
Transfer meter:
A non-converting meter supplied by the owner for the purposes of relating the volume measured by a gas measuring apparatus to a local volumetric standard.
Volume correlation:
The process by which a specific volume registered by a transfer meter or measured by a gas measuring apparatus is related to or traceable to a local volumetric standard.
Working level gas measuring apparatus:
A gas measuring apparatus intended for use in the verification, reverification and/or compliance test of gas meters.
X-bar:
The arithmetic mean of the "n" results considered.

4.0 Administrative requirements

4.1 General

4.1.1 These requirements apply immediately upon issue to all gas measuring apparatus utilizing bell prover displacement technology.

4.2 Gas measuring apparatus requirements

4.2.1 For measuring apparatus to be certified, all the requirements of this document must be evaluated and the results must meet the applicable requirements.

4.2.2 Gas measuring apparatus may be certified for testing any or all types of approved gas meters at test flow rates within the flow rate capacity of the local volumetric standard and the gas measuring apparatus.

4.2.3 The certificate issued by the designating authority must be valid for a period established by regulations for the gas measuring apparatus at the location where the calibration was completed. Any relocation or software, equipment or component replacements or modifications which affect the performance of the gas measuring apparatus require recertification of the gas measuring apparatus. The extent of the recertification is to be determined by the designating authority upon receipt of the notice referred to in clause 4.5.2(g) of this document.

4.3 Statistics

Gas measuring apparatus or related accessories intended to perform statistical calculations of average error and standard deviation of a sample of gas meters for the purposes of verification, reverification or compliance sampling must do so pursuant to the requirements of a Measurement Canada-approved statistical sampling plan for the verification or reverification of gas meters.

4.4 Certification testing

The method of certification testing must be sufficient to ensure that the gas measuring apparatus will function accurately and reliably over the conditions to which it will be subjected. These conditions include, but are not limited to, ambient air temperature, meter proving air temperature, model of meter, condition of meter, test flow rates and modes of gas measuring apparatus operation. Where it has been determined analytically or empirically that the effect of a particular condition is not significant with respect to the accuracy of a specific type of gas measuring apparatus, the method of certification testing may, with the Senior Engineer's - Gas Measurement permission, be modified to take this evidence into account. If more than one method of meter proving is to be certified, the steps in sections 5.5.3, 5.5.4 and 5.5.5 must be performed using all meter proving methods requested.

4.5 Roles and responsibilities

4.5.1 Designating authority

The designating authority is responsible for:

  1. any certification ensuing from the certification testing procedure;
  2. all calibration testing procedures and worksheet completion relevant to this document.

4.5.2 Owner

The owner is responsible for:

  1. providing a statement of intended use together with a full set of completed worksheets demonstrating that the gas measuring apparatus is fully compliant with all applicable requirements set out in this document prior to certification testing of the gas measuring apparatus by the designating authority;
  2. making all adjustments and calibrations necessary to meet the requirements;
  3. providing the transfer meters required by this document;
  4. providing the leak test apparatus required to demonstrate the ability of the gas measuring apparatus to detect the specified operational leak;
  5. using the gas measuring apparatus in the manner for which it was intended and in accordance with any conditions set out in the certificate;
  6. ensuring that the gas measuring apparatus is maintained in good repair and in the required operational order;
  7. giving the designating authority prior notification of proposed relocation, modification and/or need of repairs to the certified gas measuring apparatus. The need for recertification will be determined by the designating authority upon receipt of this notification;
  8. maintaining a log book or file which records the dates and details, including the identification of the person or persons performing accuracy checks, adjustments, maintenance, repairs and modifications to the gas measuring apparatus. The log book for each gas measuring apparatus must be made readily available to the designating authority upon request and must be retained for a period of six years;
  9. providing a stable temperature environment for the gas measuring apparatus. The prover room ambient air temperature, the meter outlet air temperature, the prover oil temperature and the gas measuring apparatus meter proving air temperature must be continuously monitored. Records of these temperatures must be maintained and reviewed prior to calibrating the gas measuring apparatus;
  10. making available to the designating authority operating instruction manuals which provide detailed information pertaining to the installation, maintenance, calibration and use of the gas measuring apparatus.

4.6 Statement of intended use of the gas measuring apparatus

4.6.1 Limitations

The owner must provide to the designating authority a detailed statement of intended use of the gas measuring apparatus. The documentation provided must be sufficient to determine the capabilities of the gas measuring apparatus, its intended uses and all installation requirements. The intended use of the gas measuring apparatus must:

  1. be within the specifications and limitations of the gas measuring apparatus published by the manufacturer; and
  2. be such that the gas measuring apparatus is capable of achieving and maintaining the required flow rates.

The minimum test volume must be as specified in Table 1 or the metric equivalent unless the owner or manufacturer can demonstrate that reducing the volume will not affect the performance of the apparatus.

Table 1: Minimum test volumes for bell provers Bell capacity Direct counting type Inferential type 2 ft3 2 ft3 0.5 ft3 5 ft3 2 ft3 0.5 ft3 10 f3 5 ft3 2 ft3

4.6.2 Statement of intended use - Details

The statement of intended use of the gas measuring apparatus must include:

  1. a full description of the gas measuring apparatus to be certified, including the manufacturer(s), operating parameters, minimum and maximum test capacities, computer software and hardware revisions, model number(s) and serial number(s);
  2. a description of each class, type or design of meter to be tested with the gas measuring apparatus;
  3. a declaration of the categories of testing for which the gas measuring apparatus is to be utilized as set out in clause 5.5.3 and modes of operation as set out in clause 5.5.5 of this document;
  4. an identification of the minimum and maximum range of test capabilities (i.e. humidity, pressure temperature, flow rate) for which certification of the gas measuring apparatus is requested; and
  5. a declaration of the method(s) of meter proving, as set out in clauses 5.5.1 and 5.5.2.

5.0 Metrological requirements

5.1 Environment

5.1.1 Temperature

5.1.1.1 The prover room ambient air temperature must be continuously maintained and monitored at ± 1 °C of the temperature chosen by the owner. The owner may change the temperature at any time during the period of the certification, but it must fall within a range of 22 °C ± 4 °C and meet the requirement of section 5.1.1.3.

5.1.1.2 The temperature of the prover room ambient air, the meter outlet air, the gas measuring apparatus bell outlet air and the prover oil must be within 0.5 °C of each other during all testing procedures and during any subsequent verification, reverification or compliance sample testing during the certification period.

5.1.1.3 Prior to and during certification testing, the prover room ambient air temperature must not vary by more than ± 1 °C and ± 0.5 °C over the previous twenty-four hour and four-hour periods, respectively.

5.2 Prover oil

5.2.1 New prover oil, when acquired for use in bell provers, must have the following properties:

  • Viscosity at 40 °C: 3.0 cSt - 7.5 cSt
  • Relative density at 15 °C: 0.82 - 0.86

5.2.2 The owner must ensure that specifications of oil purchased are met. The documentation supplied with regards to the prover oil must clearly indicate the source, brand name and batch number of the oil being used.

5.2.3 Where more than one prover has been filled with new oil from the same batch, only one representative oil sample from one of the provers needs to be tested to ensure the oil meets the specifications.

5.2.4 Prover oil must be tested annually and at the time of certification, at an approved laboratory or by the use of authorized procedures. The sample is to be drawn from the top of the prover tank.

5.2.5 The following American Society for Testing Materials (ASTM) test method specifications are applicable to testing oil for the stated properties.

Table 2: American Society for Testing Materials test methods applicable to testing oil Property ASTM test method specifications Viscosity D445: Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids (and Calculation of Dynamic Viscosity). Relative density D: Standard Test Method for Density, Relative Density, or API Gravity of Crude Petroleum and Liquid Petroleum Products by Hydrometer Method
or D: Standard Test Method for Density, Relative Density, and API Gravity of Liquids by Digital Density Meter

5.3 Mechanical requirements

The gas measuring apparatus installation and operation must be verified for compliance with the manufacturer's installation instructions and Measurement Canada requirements. Where auxiliary equipment is attached or is to be attached to the bell, the counterweight wheel or its shaft or any other movable component of the gas measuring apparatus installation not specifically mentioned, the bell balance calibrations must be performed with the auxiliary equipment attached.

5.3.1 Bell component installation

5.3.1.1 The components listed below must be checked to ensure that they are vertical. Installation of the components must be checked at least in two views spaced 90° from each other:

  1. prover tank
  2. support posts for the counterbalance wheel support frame
  3. counterbalance wheel support frame
  4. counterbalance wheel
  5. prover bell
  6. prover guide rods

5.3.1.2 The bell components must be aligned such that all components will operate in the same plane of motion throughout the entire operating range of the bell.

5.3.2 Prover oil level

5.3.2.1 The prover oil level must be checked at the start point and the maximum point of travel with the bell closed to atmosphere and the internal bell pressure adjusted to 2.00 ± 0.02 inches water column.

5.3.2.2 The sealant oil level must be measured and recorded when the bell is lowered to its maximum travel and open to atmosphere.

5.3.2.3 The prover oil level must be recorded and shown on the certificate issued by the designating authority together with the reference point.

5.3.3 Bell balance

5.3.3.1 The bell balance testing must commence at the start point used to start the proving operation and at points equal to 20, 40, 60 and 80 % of the intended range of travel of the bell and end at the maximum point of travel.

5.3.3.2 The bell must remain stationary at any and all points chosen for bell balance calibration.

5.3.4 Bell reference pressure

5.3.4.1 Upon completion of the procedure set out in clause 5.3.3, the bell pressure must be adjusted, if required.

5.3.4.2 With the bell positioned within the range of bell travel, the main counter weight must be adjusted to achieve a bell pressure of 2.00 ± 0.02 inches water column.

5.3.5 Bell static pressure

5.3.5.1 The bell static pressure calibrations must commence at the start point and at points equal to 20, 40, 60 and 80 % of the intended range of bell travel and end at the maximum point of bell travel.

5.3.5.2 The bell static pressure at all calibrated points must be equal to 2.00 ± 0.02 inches water column.

5.3.6 Bell dynamic pressure

5.3.6.1 The bell dynamic pressure calibrations must commence at the start point and at points equal to 20, 40, 60 and 80 % of the intended range of bell travel and end at the maximum point of bell travel.

5.3.6.2 The bell dynamic pressure at all calibrated points must be equal to 2.00 ± 0.02 inches water column as the bell descends at a rate not exceeding 8 inches per minute.

5.3.7 Leak test

5.3.7.1 System leak test

5.3.7.1.1 A system leak test must be conducted with all auxiliary equipment connected to the gas measuring apparatus. The piping of gas measuring apparatus must include a meter in the system leak test.

5.3.7.1.2 The system leak test must be conducted with the bell at a cardinal point near the bottom of the intended range of bell travel. The leak test time interval must be a minimum of 10 minutes during which the bell position must be monitored.

5.3.7.1.3 The position of the bell must not change during the system test.

5.3.7.1.4 When a gas measuring apparatus is equipped with an outlet control valve and the meter and associated piping up to the outlet control valve is pressurized, the entire system, including the outlet control valve, must be part of the system leak test.

5.3.7.2 Operational leak test

5.3.7.2.1 The operational leak test procedure must be incorporated into the use of the gas measuring apparatus process and must be tested for both leak detection capability and repeatability.

5.3.7.2.2 Operational leak test procedures must be capable of detecting a leak of 0.25 cubic feet per hour at 2 inches of water column or greater using a leak test duration and applied pressure/vacuum designated by the owner. The operational leak test must be initiated three consecutive times to verify the reliability and repeatability of the process.

5.3.7.2.3 The owner must provide the leak test apparatus, calibrated to the local volumetric standard or another certified reference standard, for the purpose of the operational leak test.

5.3.8 Flow rate tests

5.3.8.1 The flow rate setting mechanism of the gas measuring apparatus must be tested at both the high and low load verification test points for each meter listed in the statement of intended use.

5.3.8.2 The flow rate mechanism must be capable of setting the flow rates to within the specifications of the nominal high load and low load verification test points for each meter listed in the statement of intended use.

5.3.8.3 The flow rate setting mechanism must be tested on the gas measuring apparatus using:

  1. transfer meters of known accuracy, and/or
  2. production meters of known accuracy,
  3. designated transfer meters representing flow rates between 10 % to 150 % of rated air capacity,
  4. selected meter(s) that may use both metric and imperial units of measurement with identical flow rates.

5.3.8.4 In the case of an adjustable flow rate mechanism, the flow rate mechanism calibration must be repeated six times at the high load flow rate and six times at the low load flow rate for each meter. In the case of fixed flow rate caps, the flow rate tests must be repeated three times at the high load flow rate and three times at the low load flow rate.

5.3.8.5 Each calculated flow rate must be within ± 5 % of the designated low and high load flow rates for the meter being tested.

5.3.9 Register verification

If the gas measuring apparatus is equipped with a register ratio verification option, it must be verified. The register ratio option must be verified by utilizing both a correct and an incorrect model of register representing a metric and imperial meter designated in the statement of intended use to ensure that the system is capable of accurately detecting the correct register ratio.

5.4 Meter classifications and transfer meters

5.4.1 Meter classifications

Meters shown in the statement of intended use must be grouped according to either meter class or meter classification, depending on the method of counting used by the gas measuring apparatus. A transfer meter is chosen to represent each meter class or meter classification.

5.4.2 Transfer meters

5.4.2.1 Transfer meters representative of meters in the various meter classes or meter classifications must be used to determine the percent error of the gas measuring apparatus by comparison to the local volumetric standard.

5.4.2.2 Transfer meters must be non-converting positive displacement gas meters.

5.4.2.3 Each transfer meter must be calibrated to possess an error within the range of -2 % and -3 % at low and high load flow rates, and a maximum difference between the low load error and the high load error (spread) of 0.5 or less.

5.4.2.4 Transfer meters must be acclimatized in the area of the gas measuring apparatus for a minimum period of four hours.

5.4.2.5 It is the responsibility of the owner to ensure that selected transfer meters are proven repeatable prior to use as transfer meters. The suggested method is as follows:

  1. Exercise potential transfer meters for a minimum of five minutes at a flow rate not exceeding 50 % of the rated air capacity.
  2. Run the meter six times at both the low and high load flow rates on the local volumetric standard to determine the meter error.

The selected meter is considered acceptable for use as a transfer meter provided that the percent error of each of the runs at the specified test flow rate is within ± 0.2 of the X-bar of percent errors for all six runs (see Table 3 for example).

Table 3: Determination of meter repeatability Run #1 Run #2 Run #3 Run #4 Run #5 Run #6 X-bar
6 runs Acceptable
limits 0.5 0.6 0.4 0.6 0.5 0.5 0.52 0.52 ± 0.2

5.4.2.6 The flow rate of the local volumetric standard must be set to within ± 2 % of Qmax of the specified high and low load test points for the transfer meter to be tested. For example, the high load flow rate of a meter with a rated capacity of 180 cubic feet per hour would be within 257.4 cubic feet per hour to 264.6 cubic feet per hour and the low load flow rate would be 77.4 cubic feet per hour to 84.6 cubic feet per hour.

5.4.2.7 The flow rate of the gas measuring apparatus must be set to 145 % ± 5 % and 45 % ± 5 % of the badged flow rate of the transfer meter to be tested. The flow rate of the gas measuring apparatus must be set to within ± 5 % of Qmax of the specified high and low load test points for the transfer meter to be tested. For example, the high load flow rate of a meter with a rated capacity of 180 cubic feet per hour would be within 252 cubic feet per hour to 270 cubic feet per hour, and the low load flow rate would be 72 cubic feet per hour to 90 cubic feet per hour.

5.5 Volume correlations

5.5.1 Direct counting gas measuring apparatus

In order to test a direct counting gas measuring apparatus for the purpose of certification or recertification, a transfer meter must be chosen from each meter class listed in the statement of intended use to act as representative of the class.

5.5.2 Inferential gas measuring apparatus

In order to test an inferential gas measuring apparatus for the purpose of certification, a transfer meter of each meter classification listed in the statement of intended use must be tested. In the case of a scheduled recertification of an inferential gas measuring apparatus, a transfer meter representative of each meter class listed in the statement of intended use is chosen.

5.5.3 Correlations

5.5.3.1 Volume correlation must be made to the local volumetric standard to determine whether the gas measuring apparatus may be certified for:

  1. verification,
  2. reverification, and/or
  3. compliance sampling.

5.5.3.2 Volume correlations must be conducted at the low and high load flow rates of each meter being tested.

5.5.3.3 Volume correlations must be conducted with the gas measuring apparatus in the non-converting mode.

5.5.3.4 Testing of a gas measuring apparatus using a transfer meter must be completed on the same day that the transfer meter acceptability and proof errors were established with the local volumetric standard.

5.5.3.5 Each transfer meter must be proven six times with the gas measuring apparatus at both the low and high load flow rates of that transfer meter.

5.5.3.6 The percent error for each of the six runs must be within ± 0.2 of the X-bar of the percent errors of the meter as determined with the local volumetric standard at each flow rate.

5.5.3.7 The requirements of sections 5.5.3, 5.5.4 and 5.5.5 must be satisfied for each method of meter proving, as designated by the owner.

5.5.4 Maximum error detection

5.5.4.1 Volume correlations to determine the maximum detectable error must be completed with non-converting transfer meters of any one meter class, type or design set out in the statement of intended use. Transfer meters must be adjusted by the owner to register the following errors:

  1. for the purposes set out in clause 5.5.3.1(a) and/or (b): + 2.5 % ± 0.5 %;
  2. for the purposes set out in clause 5.5.3.1(c) only or in addition to 5.5.3.1(a) and/or (b): + 9.0 % ± 0.5 % and - 9.0 % ± 0.5 %.

5.5.4.2 The transfer meters must be run six times with the gas measuring apparatus at the high load flow rates. The X-bar of the errors of the six runs must be used to determine compliance for maximum error detection. The gas measuring apparatus must be placed in non-converting mode.

Are you interested in learning more about Bell-Prover Gas Meter Test Bench? Contact us today to secure an expert consultation!

5.5.4.3 The percent error of each transfer meter test, as determined with the gas measuring apparatus, must be within ± 0.2 of the X-bar of the percent errors as determined on the local volumetric standard.

5.5.5 Additional modes of operation

  1. Where the owner's statement of intended use includes the operation of the gas measuring apparatus in different modes of operation, each mode of operation must be tested using one transfer meter.
  2. The transfer meter must be proven six times with the gas measuring apparatus at the high load flow rate. The X-bar of the percent error of these six runs must be used to determine compliance for each additional mode of operation.
  3. The percent error of each transfer meter test, as determined with the gas measuring apparatus, must be within ± 0.2 of the X-bar of the percent errors determined with the local volumetric standard.

5.5.5.1 Temperature differential mode correlations

  1. The gas measuring apparatus must be switched to and tested in the temperature differential mode. This mode is used for correlations and the verification/reverification of non-converting diaphragm gas meters. The temperature differential mode corrects for the difference in flowing air temperature between the bell outlet and meter outlet and applies a correction factor to determine the true meter relative error.
  2. During the test, the temperature sensors for the bell outlet air temperature and the meter outlet air temperature must be made to differ by a temperature of 1.0 °C ± 0.5 °C by placing the meter outlet air temperature sensor in a temperature-controlled bath.
  3. With the temperature difference maintained, test the transfer meter six times at high load flow rate and calculate the X-bar.
  4. The high load flow rate, as indicated on the gas measuring apparatus, must be within ± 2 % of the high flow rate as indicated on the local volumetric standard. Calibration will be required if the indicated errors are outside requirements.
  5. The resulting X-bar is adjusted by calculation to compensate for the artificial temperature difference. See procedures section for applicable formula.
  6. The calculated X-bar of the errors, as determined by the gas measuring apparatus, must be within ± 0.3 of the X-bar of the percent errors of the transfer meter determined by the local volumetric standard.

5.5.5.2 Temperature converting mode correlations

  1. The gas measuring apparatus must be switched to and tested in the temperature converting mode.
  2. The resulting meter errors are adjusted by calculation to compensate for the correction applied by the gas measuring apparatus.
  3. The X-bar of the errors, determined by the gas measuring apparatus, must be within ± 0.3 of the X-bar of the percent errors of the transfer meter determined by the local volumetric standard.

6.0 Technical requirements

6.1 Use requirements

6.1.1 Weekly correlation - Transfer meter / local volumetric standard

6.1.1.1 The volume correlation of the transfer meters to the local volumetric standard must be performed:

  1. each week prior to the use of the gas measuring apparatus,
  2. using a non-converting transfer meter having an error of - 2.5 % ± 0.5 % at low and high load flow rates.
    1. using transfer meters which have been acclimatized for a minimum of four hours.
    2. using transfer meters which have been constantly exercised at a rate equal to, or less than 15 % of the badged flow rate.

6.1.1.2 The transfer meters used for the weekly volume correlation must be representative of the metric or imperial meter classifications of those meters which are to be verified, reverified or compliance tested that week.

6.1.1.3 The transfer meter must be run six times with the local volumetric standard at both the low and high load flow rates. The flow rate of the local volumetric standard must be set to 145 % ± 2 % and 45 % ± 2.0 % of the badged air flow rate of the transfer meter to be tested. The X-bar of the percent errors of these runs must be used to determine the average true errors. These values must be utilized during correlation of the gas measuring apparatus over the course of the next week.

6.1.1.4 Transfer meter performance must be tracked to ensure reliability and repeatability. Weekly errors greater than ± 0.2 for either the high or low load flow rate from the previous correlation to the local volumetric standard must be investigated and noted in the designated prover log book.

6.1.2 Daily correlation - Transfer meter / gas measuring apparatus

6.1.2.1 Volume correlation of the gas measuring apparatus must be performed:

  1. each day prior to the use of the gas measuring apparatus;
  2. using designated non-converting transfer meters from 6.1.1;
  3. using transfer meters which have been acclimatized for a minimum of four hours;
  4. with the gas measuring apparatus in the
    1. temperature differential mode if non-converting meters are to be verified or reverified; and/or
    2. temperature converting mode if temperature converting meters are to be verified or reverified.

6.1.2.2 For direct counting gas measuring apparatus, the transfer meters used must be representative of the metric or imperial meter class of those meters which are to be verified or reverified that day. For inferential gas measuring apparatus, the transfer meters used must be representative of the metric or imperial meter classification of those meters which are to be verified or reverified that day. Meters of any other classification must not be processed until the transfer meter representing that meter classification has been subjected to the daily volume correlation process. Daily volume correlations need not be performed if gas meters are not to be verified or reverified during that day.

6.1.2.3 Transfer meters must be run three times with the gas measuring apparatus at both the high load flow rate and low load flow rate. The low and high load flow rates of the gas measuring apparatus must be set to within 145 ± 5 % and 45 ± 5 % of the badged air rate of the transfer meter to be tested. The X-bars of the percent errors of these three runs must be used to determine the average true error, which must be within ± 0.2 of the percent error as established against the local volumetric standard within the previous one-week period.

6.1.2.4 Where the ± 0.2% allowable error tolerance has been exceeded, the steps below must be followed until the deficiency is resolved:

  1. repeat the correlation process shown in 6.1.2.3;
  2. repeat the weekly correlation process, pursuant to clause 6.1.1;
  3. perform a complete diagnostic analysis/check to ensure the integrity of the gas measuring apparatus; and/or
  4. where the ± 0.2 % allowable error tolerance is still not complied with, remove the gas measuring apparatus from service and initiate a nonconformance.

6.1.3 Operational leak detection

6.1.3.1 An operational leak detection sequence must be utilized prior to the final test sequence for all verification, reverification and compliance testing procedures.

6.1.3.2 The duration of the operational leak test must be as determined by the owner and specified for the test shown in section 5.3.7.2.

6.1.4 Prover oil

The owner of the gas measuring apparatus must have a sample of the prover oil tested annually and at the time of recertification for viscosity and relative density in accordance to the test methods requirement in section 5.2.5. Certificates of analysis must be made available to the designating authority upon request.

6.1.5 Temperature

6.1.5.1 The prover room ambient air temperature must be continuously maintained and monitored at ± 1 °C of the temperature chosen by the owner. The chosen temperature may be changed by the owner at any time during the period of the certification, but must fall within a range of 22 °C ± 4 °C and meet the requirement of section 5.1.1.3.

6.1.5.2 The prover room ambient air temperature, the meter outlet air, the gas measuring apparatus bell outlet air temperature and the prover oil temperature must be within 0.5 °C of each other during all testing procedures and during any subsequent verification, reverification or compliance sample testing during the certification period.

6.1.5.3 Prior to and during all verification, reverification or compliance sample testing, the prover room ambient air temperature must not vary by more than ± 1 °C and 0.5 °C over the previous twenty-four hour and four-hour periods, respectively.

6.1.5.4 Temperature records must be retained for a time period of not less than three years.

6.1.6 Maintenance

6.1.6.1 The owner must perform routine maintenance as specified in the manufacturer's and owner's manuals. As a minimum, the maintenance and/or calibration of components and sensors must be performed annually.

6.1.6.2 The calibration of pressure, temperature and other sensors must be performed with a traceable standard.

6.1.6.3 Records of maintenance and calibrations must be maintained as part of the prover log book as per section 4.5.2(h).

G-16—Recognition of gas meter test results from test facilities and ...

Category: Gas
Issue date: -02-04
Effective date: -02-04
Revision number: 11
Supersedes: G-16 (rev. 10)

Table of contents

  • 1.0 Purpose
  • 2.0 Scope
  • 3.0 Limitations
  • 4.0 Conditions and requirements
    • 4.1 Evaluation of test capacities and ranges
    • 4.2 Processes and limitations
    • 4.3 Evaluation criteria
      • 4.3.1 Traceability
      • 4.3.2 Measurement uncertainty
      • 4.3.3 Proficiency testing and monitoring of results
      • 4.3.4 Status indicators
      • 4.3.5 Test reports
      • 4.3.6 Design and construction of test equipment
      • 4.3.7 Technical procedures
  • 5.0 Recognition review
    • 5.1 List of recognized test facilities and laboratories
  • 6.0 Revisions
  • Appendix A: Recognized national measurement institutes (NMI)
    • Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany
    • VSL, Delft, The Netherlands
  • Appendix B: BIPM recognized test facilities and laboratories
    • Euroloop, Rotterdam, The Netherlands
    • FORCE Technologies, Vejen, Denmark
    • PIGSAR facility (Ruhrgas), Dorsten, Germany
    • TransCanada Calibrations (TCC), Île-des-Chênes, Manitoba, Canada
  • Appendix C: Recognized test facilities and laboratories accredited under another institute or accreditation program
    • Colorado Engineering Experiment Station Inc. (CEESI), Garner, Iowa, USA
    • Colorado Engineering Experiment Station Inc. (CEESI), Nunn, Colorado, USA
    • Colorado Engineering Experiment Station Inc. (CEESI), Roselle, Illinois, USA
    • DNV GL, Groningen, The Netherlands
    • Euroloop, Utrecht, The Netherlands
    • FortisBC Energy Inc., Triple Point Turbine Meter Calibration Facility, Penticton, B.C., Canada
    • NMi Certin B.V, Delft, The Netherlands
    • VSL, Delft, The Netherlands
  • Appendix D: Recognized test facilities and laboratories accredited under some other accreditation program
    • Dresser Inc., Houston, Texas, USA
  • Appendix E: Test facilities and laboratories whose recognition is under evaluation by Measurement Canada
    • Sensus USA Inc., DuBois, Pennsylvania, USA

1.0 Purpose

This bulletin describes the requirements and conditions for the recognition by Measurement Canada of gas meter test results from test facilities and laboratories. Test results from recognized facilities and laboratories may be used by Measurement Canada for the purposes of meter calibration, type approval and measurement dispute investigation or by a Measurement Canada accredited organization for the purposes of verification, reverification and sealing of meters.

2.0 Scope

This bulletin applies to test facilities and laboratories who want to have their test results recognized by Measurement Canada for the purposes described above or by Measurement Canada's Engineering and Laboratory Services for the purpose of generating certificates of calibration for reference meters.

3.0 Limitations

This recognition does not delegate or confer any legal authority upon the operator of the test facility or laboratory to certify measuring apparatus or to verify meters pursuant to the Electricity and Gas Inspection Act (EGIA).

To satisfy the requirements of the EGIA, the use of test results from recognized test facilities and laboratories is limited to Measurement Canada and accredited meter verifiers. In addition, recognition of test results is limited to the test capacities listed for each facility or laboratory in this bulletin's appendices.

4.0 Conditions and requirements

4.1 Evaluation of test capacities and ranges

A facility or laboratory seeking recognition of its test results must provide to Measurement Canada for evaluation purposes the required information listed in Table 1.

Table 1: Summary of required information Required information Explanation and/or examples Name and contact information Name, address, contact person's number and address Scope Pipeline meters used for the trade of natural gas and/or
calibration of reference meters for measuring apparatus Type of meter(s) under test (MUT)

Examples:

  • Rotary, turbine, ultrasonic, vortex, Coriolis mass flow, differential pressure producers
Flow range Minimum to maximum – natural gas or an alternative test medium Pressure
  • Pressure range – natural gas
  • Atmospheric – air
Temperature (test medium)
  • 5 °C to +25 °C (seasonal ambient) – natural gas
  • 20±5 °C (controlled) – air
Size of meters
  • Pipe size (diameter): NPS (or DN)
  • Flange class: ANSI rating
Relative expanded uncertainty (k = 2, 95% confidence interval)

Include ranges of flow and run designation, if applicable, and indicate the test medium and application.

Examples:

  • Pipeline meters – natural gas
  • Reference meters – air

Test medium

Examples:

  • Pipeline meters – natural gas
  • Reference meters – air or alternative test medium

Existing accreditations and reference numbers

Examples:

  • ISO/IEC , ISO , CLAS certificate number

Reference agency or NMI

Examples:

  • National Research Council Canada (NRC)
  • Measurement Canada
Supporting documentation All the listed documents and publications to be provided to Measurement Canada to support the request for recognition.

4.2 Processes and limitations

Measurement Canada's Engineering and Laboratory Services perform a technical evaluation in order to assess the test facility or laboratory's ability to provide the test results to the values stated in Table 1.

The evaluation method employed depends on the level of the facility or laboratory on the international metrological traceability chain and current certifications or accreditation. Upon recognition, the facility or laboratory is listed in one of the four appendices found in this bulletin, as described below.

Appendix A lists recognized national measurement institutes (NMIs) who are signatories to the Bureau international des poids et mesures (BIPM) Mutual recognition of national measurement standards and of calibration and measurement certificates issued by national metrology institutes () and whose calibration and measurement capabilities are listed in Appendix C of the BIPM mutual recognition agreement and evaluated by participating in the International Key Comparison. No additional review of these institutes outside of the key comparison is conducted. These institutes, by way of their inclusion in the BIPM mutual recognition agreement, are considered to have met the requirements of this bulletin.

Appendix B lists test facilities and laboratories that have been designated by a country whose NMI is listed in Appendix A. These test facilities and laboratories are supported, but not owned by the designating country. As these NMIs are signatories to the BIPM mutual recognition agreement and included in Appendix C of the agreement, no additional review is conducted by Measurement Canada as they are considered to have met the requirements of this bulletin.

Appendix C lists recognized test facilities and laboratories that have been accredited to ISO by an organization who is recognized by the International Laboratory Accreditation Cooperation (ILAC), to which the Standards Council of Canada (SCC) is a signatory member. This appendix also lists laboratories that have been accredited under the Calibration Laboratory Assessment Service (CLAS) in Canada and the National Voluntary Laboratory Accreditation Program (NVLAP) in the United States. A documentation review and on-site technical review of these facilities and laboratories are conducted by Measurement Canada to ensure they meet the requirements of this bulletin.

Appendix D lists recognized test facilities and laboratories that have been accredited under an accreditation program other than the ones mentioned in appendices A, B and C. An in-depth technical review of these facilities and laboratories is conducted by Measurement Canada to ensure they meet the requirements of this bulletin.

Appendix E lists test facilities and laboratories currently under evaluation by Measurement Canada for initial recognition or renewal of their existing recognition. Unless otherwise indicated, Measurement Canada will continue to accept test results from the recognized facilities and laboratories listed in this appendix.

4.3 Evaluation criteria

Measurement Canada's evaluation process focuses on the criteria stated in subsections 4.3.1 to 4.3.7 of this bulletin. It is therefore recommended that the documentation provided to Measurement Canada be organized in relation to these subsections.

Test facilities and laboratories must provide any technical documentation relating to sections 6.3 to 6.5 and 7.3 to 7.8 of ISO/IEC : General Requirements for the Competence of Testing and Calibration Laboratories and must reference the related sections of that standard in their documentation.

4.3.1 Traceability

Test facilities and laboratories must submit sufficient information to demonstrate traceability to a relevant national standard.

A certificate of accreditation to ISO/IEC by an accrediting body recognized by the NRC or the Standards Council of Canada (SCC) is considered as proof of compliance.

4.3.2 Measurement uncertainty

Test facilities and laboratories must state the overall measurement uncertainty of their test results and the method used in its determination. The method used must be as prescribed in the ISO Guide for the Expression of Uncertainty (GUM).

If recognition of test results is for the purpose of trade meter calibration, the expanded uncertainty must be less than 0.33% of the applicable device tolerance at the 95% confidence interval in order for the facility or laboratory to be recognized by Measurement Canada. If recognition is for the purpose of calibration of reference meters, a further reduced uncertainty value will be expected. Note that no indication of the value of uncertainty stated by any test facility or laboratory is provided in this bulletin's appendices.

4.3.3 Proficiency testing and monitoring of results

In order to assess the proficiency of test facilities and laboratories and ensure their continued compliance with the requirements of this bulletin, they must use monitoring processes such as:

  • participating in intercomparisons of meter calibrations with one or more laboratories or test facilities recognized by Measurement Canada,
  • monitoring the performance of reference standards,
  • using check or monitoring standards.

The results of these monitoring processes must be provided to Measurement Canada upon request.

Measurement Canada may require proficiency testing to assess test facilities and laboratories.

4.3.4 Status indicators

A recognized test facility or laboratory must affix a status indicator on meters that have been calibrated and are presented to Measurement Canada for verification or reverification or on meters that are the subject of a dispute. In instances where the meter can be sealed, the facility or laboratory must affix a seal on the meter in order to prevent unauthorized access to the meter's adjustments. The facility or laboratory must be identified on the affixed seal.

Meters tested or calibrated at a recognized test facility or laboratory must be verified and sealed by Measurement Canada or an accredited meter verifier, prior to being placed into service.

4.3.5 Test reports

A recognized test facility or laboratory must include the following information in its test reports:

  • a title (e.g. Test Report),
  • their name and address and the location where the tests were carried out, if different from the facility or laboratory's address,
  • a unique identifier for the test report (such as a document number) that is included on each page in order to ensure that the page is recognized as part of the test report,
  • the name and address of the organization who owns the meter and requested the testing,
  • the description and identifier (such as a serial number) of the meter that was tested, including the serial numbers of all controlled components, where required,
  • the date(s) on which the testing was performed,
  • the test or calibration results with the units of measurement, including all details necessary to demonstrate that Measurement Canada's metrological performance requirements have been assessed,
  • identification of the test procedures that were followed and any installed meter accessories or influencing components, as applicable,
  • the stated uncertainty at the 95% confidence interval associated with the test results,
  • an indication of any factors not included in the estimate of uncertainty.

4.3.6 Design and construction of test equipment

The test facility or laboratory must submit information related to the design and construction of the test equipment, process and instrumentation drawings, and instrumentation specifications for all the equipment used to obtain meter test results.

4.3.7 Technical procedures

The test facility or laboratory must submit copies of the test procedures used to calibrate meters. Note that a certificate of accreditation to ISO/IEC by an accrediting body recognized by the NRC or SCC is considered as proof of compliance.

5.0 Recognition review

Measurement Canada monitors the performance of recognized test facilities and laboratories and determines whether they should continue to be recognized based upon the results of the review. Recognized test facilities and laboratories are reevaluated every four (4) years. Measurement Canada may perform on-site audits as part of the evaluation and monitoring of recognized test facilities and laboratories.

Measurement Canada's recognition process and requirements are periodically evaluated for effectiveness and suitability, and may be modified or discontinued as necessary. Recognized test facilities and laboratories must be prepared to update their processes and procedures to meet any new or modified administrative or technical requirements. They may also be required to meet certain quality system requirements as Measurement Canada adjusts its recognition program.

5.1 List of recognized test facilities and laboratories

Detailed listings of the test facilities and laboratories that have been recognized by Measurement Canada are provided in this bulletin's appendices.

6.0 Revisions

The purpose of revision 11 was to:

  • update the Colorado Engineering Experiment Station Inc.(CEESI) capabilities with respect to flow range, pressure range and pipe size,
  • remove the Rosselle facility from the bulletin, as requested by the facility.

The purpose of revision 10 was to update the flow range and size of meters shown for the TransCanada Calibrations (TCC) facility.

The purpose of revision 9 was to:

  • correct the temperature range for VSL apparatuses NE04 and NE05,
  • add recognition of the VSL travelling reference meter method,
  • add recognition of the DNV GL test facility in Groningen, Netherlands.

The purpose of revision 8 was to add recognition of the NMi Certin B.V. test facility.

The purpose of revision 7 was to update the Colorado Engineering Experiment Station Inc.(CEESI) capabilities.

The purpose of revision 6 was to:

  • add recognition of the FORCE and Euroloop test facilities,
  • remove the Gasunie Research facility,
  • update all other recognized laboratory capabilities where needed,
  • update the ISO references to the most current version,
  • make formatting changes.

The purpose of revision 5 was to:

  • recognize name changes for the NMI and Terasen Gas,
  • make minor editorial changes in claimed laboratory capabilities,
  • clarify the purpose of the bulletin as it relates to accredited meter verifiers.

The purpose of revision 4 was to:

  • remove the former restriction on the use of test results from recognized test facilities by organizations accredited by Measurement Canada,
  • correct typographical errors in the rated maximum flow rate shown for various test facilities.

The purpose of revision 3 was to:

  • reformat the bulletin,
  • add information relating to the assessment process,
  • update and expand the information concerning previously recognized test facilities,
  • lapse the recognition of the TransCanada Pipelines Ltd. (TCPL) test facility,
  • add recognition of the Physikalisch-Technische Bundesanstalt (PTB), Terasen Gas Inc. and the Dresser, Inc. test facilities.

The purpose of revision 2 was to:

  • add recognition of the TransCanada Pipelines Ltd. (TCPL) test facility,
  • change "Equimeter Inc." to "Invensys Energy Metering".

The purpose of revision 1 was to add recognition of the TransCanada Calibrations Ltd. (TCC) test facility.

If you want to learn more, please visit our website Water Meter Test Bench.

Appendix A: Recognized national measurement institutes (NMI)

Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany

Scope
Calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC
Reference agency or NMI
Physikalisch-Technische Bundesanstalt (PTB)
BIPM Key Comparison Database, Appendix C – Calibration and Measurement Capabilities, Mass and Related Quantities (DE35, DE36, DE37, DE43, DE44)
Apparatus
DE35
Type of MUT
Rotary, turbine, ultrasonic, vortex and orifice meters
Flow Range(s)
  • 0.1 to m3/h
  • 0.12 to kg/h
Pressure
Atmospheric
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS ⅜" to 16" (DN 10 to DN 400)
Test medium
Air
Apparatus
DE36
Type of MUT
Turbine and rotary meters
Flow range
  • 200 to 24,000 m3/h
  • 240 to 28,800 kg/h
Pressure
Atmospheric
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 4" to 24" (DN 100 to DN 600)
Test medium
Air
Apparatus
DE37
Type of MUT
Turbine and rotary meters
Flow range
  • 3 to 480 m3/h
  • 200 to 22,000 kg/h
Pressure
0.8 to 5.6 MPa
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 2" to 6" (DN 50 to DN 150)
Test medium
Natural gas
Apparatus
DE43
Type of MUT
Turbine, rotary and ultrasonic meters
Flow range
  • 0.015 to 100 m3/h
  • 0.03 to 800 kg/h
Pressure
0.12 to 0.8 MPa
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 1" to 4" (DN 25 to DN 100)
Test medium
Air
Apparatus
DE44
Type of MUT
Turbine, rotary and ultrasonic meters
Flow range
  • 0.015 to 250 m3/h
  • 0.03 to 350 kg/h
Pressure
0.1 to 0.5 MPa
Temperature
  • 20±5 °C (ambient)
  • 45 °C to 80 °C (flowing)
Size of meters
Pipe size: NPS 1" to 4" (DN 25 to DN 100)
Test medium
Air

VSL, Delft, The Netherlands

Scope
Calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC
Reference agency or NMI
  • VSL
  • BIPM Key Comparison Database, Appendix C – Calibration and Measurement Capabilities, Mass and Related Quantities (NE04, NE05, NE08)
Apparatus
Bell prover 3 (NE04)
Type of MUT
Rotary, turbine meters
Flow range
1 to 400 m3/h
Pressure
Ambient air
Temperature
20±0.5 °C (controlled)
Size of meters
Pipe size: NPS ⅜" to 4" (DN 10 to DN 100)
Test medium
Air
Apparatus
Large test bench (NE05)
Type of MUT
Rotary, turbine, ultrasonic meters
Flow range
15 to 15,000 m3/h
Pressure
Ambient air
Temperature
20±0.5 °C (controlled)
Size of meters
Pipe size: NPS 3" to 24" (DN 80 to DN 600)
Test medium
Air
Apparatus
Gas oil piston prover (NE08)
Type of MUT
Rotary meters
Flow range
5 to 230 m3/h
Pressure
0.1 to 6.1 MPa
Temperature
20±5 °C (controlled)
Size of meters
Pipe size: NPS 4" (DN 100)
Test medium
Natural gas

Appendix B: BIPM recognized test facilities and laboratories

Euroloop, Rotterdam, The Netherlands

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC : (RvA K 161)
Reference agency or NMI
  • Measurements traceable to VSL, The Netherlands
  • BIPM Key Comparison Database, Appendix C – Calibration and Measurement Capabilities, Mass and Related Quantities (VSL/NE06)
Type of MUT
Rotary, turbine, ultrasonic and Coriolis meters
Flow ranges
  • 20 to 30,000 m3/h
  • 200 to 1,600,000 kg/h
Pressure
0.9 to 6.3 MPa
Temperature
Partly controlled
Size of meters
Pipe size: NPS 4" to 36" (DN 100 to DN 900)
Test medium
Natural gas

FORCE Technologies, Vejen, Denmark

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC
Reference agency or NMI
  • FORCE Technologies, Denmark
  • BIPM Key Comparison Database, Appendix C – Calibration and Measurement Capabilities, Mass and Related Quantities (DK4, DK5, DK6, DK11, DK42)
Apparatus
DK4
Type of MUT
Turbine, ultrasonic, vortex and Coriolis meters
Flow range
  • 1 to 4,000 m3/h
  • 1 to 5,200 kg/h
Pressure
Atmospheric
Temperature
20±2 °C (ambient)
Size of meters
Pipe size: NPS 1" to 12" (DN 25 to DN 300)
Test medium
Air
Apparatus
DK5
Type of MUT
Turbine, ultrasonic, vortex, rotary and Coriolis meters
Flow range
65 to 25,000 m3/h
Pressure
Atmospheric
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 1" to 50" (DN 25 to DN )
Test medium
Air
Apparatus
Line 2 (DK6)
Type of MUT
Turbine, ultrasonic, vortex and rotary meters
Flow range
5 to 5,000 m3/h
Pressure
0 MPa to 0.8 MPa
Temperature
20±3 °C (ambient)
Size of meters
Pipe size: NPS 1" to 8" (DN 25 to DN 200)
Test medium
Air
Apparatus
DK11
Type of MUT
Turbine, ultrasonic, vortex, differential pressure and Coriolis meters
Flow range
5 to 32,000 m3/h
Pressure
0.3 to 6.5 MPa
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 1" to 50" (DN 25 to DN )
Test medium
Natural gas
Apparatus
Piston prover (DK42)
Type of MUT
Turbine, vortex and rotary meters
Flow range
2 to 400 m3/h
Pressure
0 MPa to 6.5 MPa
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 1" to 6" (DN 25 to DN 150)
Test medium
Air and natural gas

PIGSAR facility (Ruhrgas), Dorsten, Germany

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC
Reference agency or NMI
  • Physikalisch-Technische Bundesanstalt (PTB), Germany
  • BIPM Key Comparison Database, Appendix C – Calibration and
  • Measurement Capabilities, Mass and Related Quantities (DE38)
Type of MUT
Turbine, ultrasonic and Coriolis meters
Flow range
  • 3 to 7,200 m3/h
  • 23 to 350,000 kg/h
Pressure
1.5 to 5.6 MPa
Temperature
Partly controlled ambient
Size of meters
  • Pipe size: NPS 3" to 16" (DN 80 to DN 400)
  • Flange class: ANSI 600
Test medium
Natural gas

TransCanada Calibrations (TCC), Île-des-Chênes, Manitoba, Canada

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
  • ISO/IEC :
  • ISO
  • NRC CLAS certificate number -04
Reference agency or NMI
  • National Research Council Canada
  • Measurements traceable to VSL, The Netherlands
  • BIPM Key Comparison Database, Appendix C – Calibration and
  • Measurement Capabilities, Mass and Related Quantities (TCC/9)
Type of MUT
Volumetric flow meters
Flow range
10 to 55,000 m3/h
Pressure
6.4 ±0.5 MPa
Temperature
Partly controlled
Size of meters
  • Pipe size: NPS 1" to 42" (DN 25 to DN )
  • Flange class: ANSI 600, 900, ,
Test medium
Natural gas

Appendix C: Recognized test facilities and laboratories accredited under another institute or accreditation program

Colorado Engineering Experiment Station Inc. (CEESI), Garner, Iowa, USA

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
  • ISO/IEC
  • A2LA Certificate Number .01
Reference agency or NMI
National Institute of Standards and Technology (NIST)
Types of MUT
  • Rotary, turbine, ultrasonic, vortex and Coriolis meters, and
  • differential pressure producers
Flow range
900 to 1,330,000 acfh
Pressure
psia
Temperature
13 °C to 28 °C (seasonal ambient)
Size of meters
Pipe size: NPS 2" to 36" (DN 50 to DN 900)
Test medium
Natural gas

Colorado Engineering Experiment Station Inc. (CEESI), Nunn, Colorado, USA

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
  • ISO/IEC
  • A2LA Certificate Number .01
Reference agency or NMI
National Institute of Standards and Technology (NIST)
Type of MUT
Rotary, turbine, ultrasonic, vortex and Coriolis meters, and differential pressure producers
Flow range
0.005 to 94 lb/s
Pressure
Atmospheric to psia
Temperature
15 °C to 38 °C
Size of meters
Pipe size: NPS ¼" to 12" (DN 8 to DN 300)
Test medium
Natural gas
Type of MUT
Rotary, turbine, ultrasonic, vortex and Coriolis meters, and differential pressure producers
Flow range
0.005 to lb/min
Pressure
Atmospheric to psia
Temperature
15 °C to 38 °C
Size of meters
Pipe size: NPS ¼" to 36" (DN 8 to DN 900)
Test medium
Air

DNV GL, Groningen, The Netherlands

Scope
Meters used for the trade of natural gas
Existing accreditations
ISO/IEC : (RvA K 103)
Reference agency or NMI
Measurements traceable to Force Technologies, Denmark
Type of MUT
Rotary, turbine, ultrasonic, vortex, and Coriolis meters
Flow range
422 to 29,300 m3/h
Pressure
0.9 to 4.1 MPa (abs.)
Temperature
18 °C to 30 °C
Size of meters
Pipe size: NPS 1" to NPS 8" (DN 25 to DN 200)
Test medium
Natural gas

Euroloop, Utrecht, The Netherlands

Scope
Meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC : (RvA K 161)
Reference agency or NMI
Measurements traceable to VSL, The Netherlands
Type of MUT
Turbine and ultrasonic meters
Flow range
5 to 10,000 m3/h
Pressure
0.85 to 0.95 MPa (abs.)
Temperature
Partly controlled
Size of meters
Pipe size: NPS 6" to NPS 16" (DN 150 to DN 400)
Test medium
Natural gas

FortisBC Energy Inc., Triple Point Turbine Meter Calibration Facility, Penticton, B.C., Canada

Scope
Pipeline meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
  • Measurement Canada S-A-01 certificate number A
  • ISO/IEC
  • Accredited laboratory number 778
  • NRC CLAS certificate number -01
Reference agency or NMI
Measurements traceable to FORCE Technologies, Denmark
Type of MUT
Turbine and Coriolis meters
Flow range
  • 10 to 6,500 m3/h
  • 350 to 230,000 acfh
  • 1,600 to 18,250 lb/hr
Pressure
  • 20 to 1,654 kPa – absolute
  • 3 to 240 psia
Temperature
5.0 °C to 35 °C (41 °F to 95 °F) – controlled
Size of meters
  • Pipe size: NPS 2" to 12" (DN 50 to DN 300)
  • Flange class : ANSI 150, 300 and 600
Test medium
Carbon dioxide gas (CO2)

NMi Certin B.V, Delft, The Netherlands

Scope
Type-approval evaluation of gas meters and measuring systems for gaseous fuels
Existing accreditations
  • ISO : RvA I122
  • ISO : RvA L029
Reference agency or NMI
Measurements traceable to VSL, The Netherlands
Type of MUT
Diaphragm, rotary, turbine, thermal mass, ultrasonic, Coriolis and vortex meters, electronic volume conversion devices (EVCD), calorific value determining devices (CVDD) and gas chromatographs
Flow range
0.4 to m3/h
Pressure
Atmospheric to 100 mbar(g)
Temperature
-32 °C to 62 °C
Size of meters
Pipe size: NPS 3/8" to 1½" (DN 10 to DN 40)
Test medium
Air
Type of MUT
Diaphragm, rotary, turbine, thermal mass, ultrasonic, Coriolis and vortex meters, EVCD, CVDD and gas chromatographs
Flow range
0.016 to 10 m3/h
Pressure
35 to 65 mbar(g)
Temperature
-32 °C to 62 °C
Size of meters
Pipe size: NPS 3/8" to 1½" (DN 10 to DN 40)
Test medium
Pure methane or H, L and E gases according to EN 437

VSL, Delft, The Netherlands

Scope
Calibration of reference meters for measuring apparatus
Existing accreditations
ISO/IEC (RvA scope number K999)
Reference agency or NMI
Measurements traceable to VSL, The Netherlands
Method
On-site calibration via travelling reference meters
Type of MUT
Rotary, turbine, ultrasonic and vortex meters
Flow range
0.1 to 10,000 m3/h
Pressure
Ambient air
Temperature
20±5 °C (ambient)
Size of meters
Pipe size: NPS 2" to NPS 20" (DN 50 to DN 500)
Test medium
Air

Appendix D: Recognized test facilities and laboratories accredited under some other accreditation program

Dresser Inc., Houston, Texas, USA

Scope
Calibration of reference meters for measuring apparatus
Existing accreditations
Nederlands Meetinstituut (NMI) traceable to project number ()
Reference agency or NMI
Nederlands Meetinstituut (NMI) and NMI Certin B.V. (), The Netherlands
Apparatus
Bell prover (50 ft3, 20 ft3 and 10 ft3)
Type of MUT
Rotary reference meters used in Dresser Model 5 transfer provers
Flow range
35 to 6,600 ft3/h
Pressure
Atmospheric
Temperature
21±1 °C
Size of meters
Pipe size: NPS 1.5" to 4" (DN 40 to DN 100)
Flange class: ANSI 150, 300
Test medium
Air
Apparatus
Piston prover ( ft3)
Type of MUT
Rotary reference meters used in Dresser Model 5 transfer provers
Flow range
to 87,250 ft3/h
Pressure
Atmospheric
Temperature
21±1 °C
Size of meters
Pipe size: NPS 4" to 12" (DN 100 to DN 300)
Flange class: ANSI 150, 300
Test medium
Air

Appendix E: Test facilities and laboratories whose recognition is under evaluation by Measurement Canada

Sensus USA Inc., DuBois, Pennsylvania, USA

Scope
Pipeline meters used for the trade of natural gas, and calibration of reference meters for measuring apparatus
Existing accreditations
ISO :, certification number -IS7 and
ANSI/NCSL Z540-1 calibration compliance
Reference agency or NMI
National Institute of Standards and Technology (NIST)
Apparatus
50 ft3 bell prover
Type of MUT
Turbine, rotary and ultrasonic meters
Flow range
1 to 115 m3/h
Pressure
Atmospheric
Temperature
22.5±1.5 °C (controlled)
Size of meters
Pipe size: NPS 1" to 4" (DN 25 to DN 100)
Test medium
Air
Apparatus
500 ft3 bell prover
Type of MUT
Turbine meters
Flow range
0.5 to 4,000 m3/h
Pressure
Atmospheric
Temperature
22.5±1.5 °C (controlled)
Size of meters
Pipe size: NPS 2" to 12" (DN 50 to DN 300)
Test medium
Air
Apparatus
Recirculating loop and vacuum lines
Type of MUT
Turbine meters
Flow range
10 to 107,000 m3/h
Pressure
0.17 to 6.2 MPa
Temperature
27 °C to 49 °C
Size of meters
Pipe size: NPS 2" to 12" (DN 50 to DN 300)
Test medium
Air